• sensor fabrication under quality management system
  • reference and counter electrode on chip
  • up to 12 modified platinum electrodes as electro-chemical transducers
  • redundant biosensors maximize analytical reliability
  • photo-patternable nano-composite membranes
  • simultaneous measurement of glucose, lactate, glutamine, glutamate, and pyruvate
  • activity optimum at physiological dynamic range, neutral pH
  • for in vivo and ex vivo applications in whole blood and fermentation broth
  • internal volumes 0.15 - 0.4  µl
  • flow rates used range from 0.1 to 100 µl/min
  • array stability of typically at least one week at continuous operation
  • flow rate sensor on chip
  • rugged low volume fluid connection
  • micro flow-system with nanoliter internal volume
  • all plastics flexible mass producible devices
  • traceability by unique bar code label

Copyright Jobst Technologies GmbH 2010